612 Chapter7
Language Reference
SRCTKPK Source Tracking Peak
SRCTKPK Source Tracking Peak
Syntax
Figure 7-259 SRCTKPK Syntax
Description
The SRCTKPK command activates a routine that automatically adjusts
both the coarse- and fine-tracking adjustments to obtain the peak
response of the tracking generator on the spectrum-analyzer display.
Tracking peak is not necessary for resolution bandwidths greater than
or equal to 300 kHz. A thru connection should be made prior to peaking
in order to ensure accuracy.
SRCTKPK is available only with an 8560E/EC Option 002.
Example
10 OUTPUT 718;"IP;SNGLS;"
20 OUTPUT 718;"FA 300KHZ;FB 1GHZ;"
30 OUTPUT 718;"SRCPWR ON;"
40 OUTPUT 718;"SWPCPL SR;"
50 OUTPUT 718;"RB 100KHZ;"
50 OUTPUT 718;"SRCTKPK;DONE?;"
60 ENTER 718;Done
70 LOCAL 718
80 END