10-20 Programmer’s Guide
Menu Map with SCPI Commands
Menu Map for 8712ET/ES and 8714ET/ES
Table 10-5 CAL Functions, 8712ET and 8714ET (2 of 3)
KEYSTROKES SCPI COMMAND
CAL, (Transmissn) (hardkey entry)
Default Response SENS[1|2]:CORR:CSET DEF; *WAI
Response SENS[1|2]:CORR:COLL:IST OFF;METH TRAN1;
*WAI
Measure Standard SENS[1|2]:CORR:COLL STAN1;*WAI
:SENS[1|2]:CORR:COLL:SAVE;*WAI
Response & Isolation SENS[1|2]:CORR:COLL:IST OFF;METH TRAN2;
*WAI
Measure Standard - Load SENS[1|2]:CORR:COLL STAN1;*WAI
Measure Standard - Through SENS[1|2]:CORR:COLL STAN2;*WAI
Enhanced Response SENS[1|2]:CORR:COLL:IST OFF;METH TRAN3;
*WAI
Measure Standard - Open SENS[1|2]:CORR:COLL STAN1;*WAI
Measure Standard - Short SENS[1|2]:CORR:COLL STAN2;*WAI
Measure Standard - Load SENS[1|2]:CORR:COLLSTAN3;*WAI;
:SENS[1|2]:CORR:COLL:SAVE;*WAI
Measure Standard - Through SENS[1|2]:CORR:COLL STAN4;*WAI
:SENS[1|2]:CORR:COLL:SAVE;*WAI