Programmer’s Guide 11-41
SCPI Command Summary
SCPI Device Command Summary
Table 11-19 SENSe (4 of 16)
SUBSYSTEM COMMANDS FORM DESCRIPTION
SENSe[1|2]:CORRection:CLASs
[:SELect]?
CHAR Returns the current calibration
choice. Returns string from the
<item> list (below). (ES models only)
SENSe[1|2]:CORRection:CLASs
[:SELect] <item>
command
only
Selects an existing calibration from
the following <item> list:
DEFault1 Default factory one-port calibration
DEFault2 Default factory two-port calibration
REFLection3* User one-port calibration
TRANsmission1* User response calibration
TRANsmission2* User response and isolation calibration
TRANsmission3* User enhanced response calibration
TESTset1 Multi-port testset calibration
TWOPort* User two-port calibration Multi-port testset
calibration
* The requested user calibration may or may not be stored in memory, and
may or may not be valid for the current instrument settings. If the requested
user calibration is invalid, a valid user calibration will be selected, if found.
If no valid user calibration is found, the default factory calibration will be
selected. The instrument can be queried with
SENSe[1|2]:CORRection:CLASs[:SELect]? to determine the current
calibration choice.
SENSe[1|2]:CORRection:COLLect
:ABORt
command
only
Aborts the calibration that is
currently in progress.

Table of Contents

Other manuals for Agilent Technologies 8712ET

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8712ET and is the answer not in the manual?

Agilent Technologies 8712ET Specifications

General IconGeneral
BrandAgilent Technologies
Model8712ET
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals