1 Installing the GC
Installation and First Startup 17
Notes
1 Some US labs have 3 phase service resulting in 208 V at the wall receptacle. It is important
for a qualified person to measure the line voltage at the receptacle for the GC. Option 003,
208 V fast oven, uses a 220 V unit with operating range of 193 to 231 V.
2 Power line conditioners should not be used with the GC.
Power cords available
Table 2 lists the power cords available for the GC. If your power cord is incorrect, order the
cord appropriate for the country.
Fast Japan: 200 V split phase, +10 to ā10% 48-63 Hz 14.8 amps 2950 VA 15 A
Fast 220/230/240 V (2)(3) single/split phase, +10 to
ā10%
48-63 Hz 13.4 / 12.8 /
12.3 amps
2950 VA 15 A
Table 1 GC power requirements (continued)
Oven Line voltage Frequency Current Power Power outlet
current rating
Table 2 Power cords by country
Part number Country Description Wall termination Plug termination
8120-1369 Australia, New
Zealand
Power Cord, Australia/NZ, C13, 10 amp AS 3112
8120-1378 Americas Cable Assembly-Power Cord 18AWG
2.3m-LG
NEMA 5-15
8120-1689 Europe Cable Assembly-Power Cord 17AWG
2m-LG PV
CEE/7/V11
8120-2104 Switzerland Cable-Assembly-Power cord 250VAC 10A
3-C
Swiss/Denmark 1302