A Making Swagelok Connections
54 Installation and First Startup
Making Swagelok Connections
Objective
To make a tubing connection that does not leak and that can be taken apart without damaging
the fitting.
Materials needed
1/8-inch (or 1/4-inch, if used) preconditioned copper tubing
1/8-inch (or 1/4-inch, if used) Swagelok nuts
Front and back ferrules
Two 7/16-inch (for 1/8-inch nuts) or 9/16-inch (for 1/4-inch nuts) wrenches
Procedure
1 Place a Swagelok nut, back ferrule, and front ferrule on the tubing as shown in Figure 14.
Figure 14. Swagelok nuts and ferrules
2 Push the tubing into the stainless steel fitting.
3 Make sure that the front ferrule is fully engaged. Slide the Swagelok nut over the ferrule and
thread it onto the plug.
4 Push the tube fully into the plug, then withdraw it approximately 1 to 2 mm. See Figure 15.
Front ferrule
Back ferrule
Nut
Tubing
Important!
The narrow end of the
back ferrule fits into
the rear of the front
ferrule.

Table of Contents

Other manuals for Agilent Technologies 8860

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8860 and is the answer not in the manual?

Agilent Technologies 8860 Specifications

General IconGeneral
BrandAgilent Technologies
Model8860
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals