EasyManuals Logo
Home>Agilent Technologies>Laboratory Equipment>8860

Agilent Technologies 8860 Installation And First Startup

Agilent Technologies 8860
76 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #51 background imageLoading...
Page #51 background image
1 Installing the GC
Installation and First Startup 51
Step 19 When the system stabilizes, run one
injection
Perform the checkout procedure as described in the Operation Manual.
1 Enter the parameters for the checkout procedure.
If using an Agilent data system, use it to create a checkout method.
If not using a data system, enter the setpoints using the Browser Interface.
2 When the GC is ready to begin a run, the Touchscreen status shows STATUS: NORMAL
READY (alternately, when the component of the GC is not ready to begin a run, the
touchscreen shows STATUS: NOT READY), make the injection and start the run.
For a manual injection, inject the sample and press Prep Run .
3 Evaluate results by comparing the chromatogram you generated with the one in the
checkout procedure. There should be a close resemblance.

Table of Contents

Other manuals for Agilent Technologies 8860

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8860 and is the answer not in the manual?

Agilent Technologies 8860 Specifications

General IconGeneral
BrandAgilent Technologies
Model8860
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals