1 Installing the GC
Installation and First Startup 35
4 Connect the calibration gas cylinder output tubing to the hydrogen sensor module.
• Use the tubing, nuts, and ferrules provided in the hydrogen sensor kit.
• See the figure under step 1.
5 Install the calibration gas cylinder into the stand and secure using the screw.
6 Turn on the supply pressure at the pressure regulator. (You will adjust it later.)
7 Check for leaks in the fittings using a leak detection fluid. Correct any leaks.