8-43
Instrument BASIC
Programming and Using the TESTS Subsystem
880: Indicate that one of the tests have failed.
890: The goto for the end of the program.
900: End of the main program.
910: Subroutine T01-This corresponds with test #1. This subroutine illustrates how
to enter values from the Parameters, Configuration, and Specification screens.
920-930: Includes the common variables.
940: Dimension some variables that will be used to store values from the configure
screen.
960: Indicate that the first test is now active.
980: Enter the value of the first Parameter. This is the value of the first parameter on
the Parameter Screen.
990: Initialize the I_o$ string.
1000: Enter the value.
1020-1050: If there is no defined parameter this string will catch the error and return
it to the main program.
1080: Get the information for the first instrument stored on the configure screen.
1090: Initialize the I_o$ string.
1100: Enter the string.
1110: Calling_name$ now holds the string associated with the Calling Name field
on the configure screen.
1120: Model$ now holds the string associated with the Model field on the configure
screen.
1130: I1addr equals the value in the Addr field on the configure screen.
1140: Options$ now holds the string associated with the Options field on the
configure screen.
1160: Get the information for the first Specification listed on the Specification
system.
1170: Initialize the I_o$ string to null.
1180: ENTER the I_o$ string.
1190-1220: If there is no specification defined for this specification number, then an
Error will appear in the I_o$ string. If this occurs, stop the test and return the error to
the main program.