2-23
Making Measurements
Measurements
Table 3 Adjusted Limits Based on 3-Pole Filter Output RF Spectrum Due to
Modulation
Power
Control
Level
Offset from Carrier (kHZ)
0 100 200 250 400 600 to
1800
0 (43 dBm) 0 dB 0.5 dB -27.5 dB -32 dB -51 dB -70 dB
0 (39 dBm) 0 0.5 -27.5 -32 -51 -66
0 (37 dBm) 0 0.5 -27.5 -32 -51 -64
5
( 33 dBm)
0 0.5 -27.5 -32 -51 -60
Table 4 Adjusted Limits Based on 3-Pole Filter Output RF Spectrum Due to Switching
PowerLevel(dBm) Offset from Carrier (kHZ)
400 600 1200 1800
43 -3 dB -17 dB -19 dB -24 dB
41 -5 -17 -19 -24
39 -7 -17 -19 -24
37 -9 -17 -19 -24
35 -11 -17 -19 -24
33 -13 -17 -19 -24
31 -15 -19 -21 -26
29 -17 -21 -23 -28
27 -17 -22 -25 -30
25 -17 -22 -27 -32
23 -17 -22 -29 -34
21 -17 -22 -30 -36