Chapter 4 4-41
Troubleshooting
201 to 799: Device-Specific Errors
224 Option not licensed.
The selected option requires a license. Refer to the
installation procedures in the user’s guide available for
this particular option.
209 Preselector centering failed
An attempt to center the preselector failed.
704 Printer interface error
An error occurred while trying to print. Make sure the
printer is turned on and properly connected.
705 Printer Type is None
The current printer type is set to
None, so no print
operations are possible. Change the type in the
Print Setup menu and try again.
211 RBW limited to 1kHz when Span > 5MHz
In spans greater than 5 MHz, narrow (digital)
resolution bandwidths, below 1 kHz, are not available.
217 RS-232 Interface Error
An error occurred on the serial interface.
213 Span limited to 5MHz when RBW < 1kHz
In narrow (digital) resolution bandwidths, below 1 kHz,
spans greater than 5 MHz are not available.
771 Store Ref trace before turning on Normalize
A reference trace must be available for the Normalize
function to be activated. Refer to “View/Trace” on
page 6-105 where the
Normalize key function is
explained in detail.

Table of Contents

Other manuals for Agilent Technologies E4408B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E4408B and is the answer not in the manual?

Agilent Technologies E4408B Specifications

General IconGeneral
BrandAgilent Technologies
ModelE4408B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals