Page #147 background image
2-22
Assembly Replacement
A5 Sampler, A6 Frac-N, A7 Reference (Standard and Option UNR), and A8 Output
A5 Sampler, A6 Frac-N, A7 Reference (Standard and Option
UNR), and A8 Output
Tools Required
•none
Removal Procedure
Refer to Figure 2-9 for this procedure.
1. Disconnect the power cord.
2. Remove the covers from the signal generator. Refer to "Outer Instrument Cover" on page 2-5
and "Inner Instrument Cover" on page 2-7.
3. Position the signal generator with the front panel facing you.
4. Lift the retention levers and disconnect the board assembly from the A31 Motherboard.
5. Lift the board assembly out of the slot.
Replacement Procedure
Reverse the order of the removal procedure.
Perform the post-repair adjustments and performance tests that pertain to this removal
procedure.