4-2
Post-Repair Procedures
Post-Repair Procedures Matrix
Post-Repair Procedures Matrix
Table 4-1 Adjustments: Assemblies A1—A12
Adjustments Replaced Assembly
The following adjustments are listed in the order that
they should be performed for proper calibration.
A1 Keyboard
A2 Display
A2DS1 Display Backlight
A3 Power Switch
A4 Inverter
A5 Sampler
A6 Frac-N
A7 Reference (Standard)
A7 Reference (UNR)
A8 Output (Analog)
A8 Output (Digital)
A9 YIG Driver
A10 ALC
A11 Pulse/Analog Mod Gen
A12 Low Band Fast Pulse 1E6
ADC Calibration
✓ ✓ ✓ ✓ ✓ ✓ ✓ ✓
YIG-Driver Pre-Tune Calibration
✓
VCO Bias Frac-N Calibration
✓
Internal Source Calibration
✓
AM Audio Path Offset Calibration
✓ ✓
KV vs. Frequency Calibration
✓
Timebase Calibration (Non UNR/UNJ Options)
✓
FM Scale Offset Calibration
✓✓✓
FM Path Offset Calibration
✓ ✓ ✓
FM In-band Offset Calibration
✓✓✓
FM Inverting Amplifier Offset Calibration
✓
FM 1/2 Path Ratio Gain Calibration
✓✓✓
Mod Source Relative Gain Calibration
✓ ✓ ✓
FM/PM Out-of-Band Calibration
✓✓✓
FM/PM YO Frequency Compensation Calibration
✓ ✓ ✓
DC FM Calibration
✓✓✓
Low Frequency Output Calibration
✓
External Input Peak Detector Calibration
✓
ALC Dynamic Calibration
✓ ✓ ✓ ✓
Power Flatness Calibration
✓✓✓✓
Attenuator Calibration
ALC Modulator Calibration
✓✓ ✓
AM Gain Calibration
✓ ✓ ✓ ✓
LB Digital Prelevel Gain Calibration
✓
LB Digital Mixer Bias Calibration
✓
LB Digital Gain Calibration
✓
Gain Adjustment Bypass Calibration (Frequencies. ≤3.2 GHz)
✓ ✓ ✓ ✓
IQ 20 Prelevel Gain Mixer Bias, & Digital Gain Cal.
IQ Impairment Calibration
✓
Pulse Width Calibration
✓✓ ✓✓
IQ Power Floor Calibration
✓
Target IQ Power Calibration
✓
1/f Corner Calibration
✓