EasyManuals Logo

Agilent Technologies E8257C User Manual

Agilent Technologies E8257C
291 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #155 background imageLoading...
Page #155 background image
2-30
Assembly Replacement
A12 Pulse Modulator Board (Option 1E6)
A12 Pulse Modulator Board (Option 1E6)
Tools Required
T-10 driver
5/16” open ended wrench
Removal Procedure
Refer to Figure 2-13 for this procedure.
1. Disconnect the power cord.
2. Remove the covers from the signal generator. Refer to "Outer Instrument Cover" on page 2-5
and "Inner Instrument Cover" on page 2-7.
3. Position the signal generator so that the A31 Motherboard is facing you and the RF deck is on
top.
4. Using the 5/16” open ended wrench, disconnect the W16, W75, and W76 RF cables from the A12
Pulse Modulator Board.
5. Disconnect the W69 ribbon cable from the A12 Pulse Modulator Board.
6. Using the T-10 driver, remove the two screws (1) that attach the A12 Pulse Modulator Board to
the main chassis.
7. Remove the A12 Pulse Modulator Board from the RF deck.
8. Using the T-10 driver, remove the sheet metal from A12 Pulse Modulator Board.
Replacement Procedure
Reverse the order of the removal procedure.
Perform the post-repair adjustments and performance tests that pertain to this removal
procedure.

Table of Contents

Other manuals for Agilent Technologies E8257C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E8257C and is the answer not in the manual?

Agilent Technologies E8257C Specifications

General IconGeneral
BrandAgilent Technologies
ModelE8257C
CategoryPortable Generator
LanguageEnglish

Related product manuals