2-36
Assembly Replacement
A17 Baseband Generator Interface (E8267C Models Only)
A17 Baseband Generator Interface (E8267C Models Only)
Tools Required
5.0 mm hex-nut driver
9/16” hex-nut driver
needle-nose pliers
Removal Procedure
Refer to Figure 2-16 for this procedure.
1. Disconnect the power cord.
2. Remove the covers from the signal generator. Refer to "Outer Instrument Cover" on page 2-5
and "Inner Instrument Cover" on page 2-7.
3. Remove the rear panel. Refer to "Rear Panel (Standard)" on page 2-44 or
"Rear Panel (Option 1EM)" on page 2-47.
4. Disconnect the W64 ribbon cable from A17P2.
5. Disconnect the W90 ribbon cable from A17P1.
6. Using the needle-nose pliers, disconnect the W68 wire cable from A17P4.
7. Using the 9/16” hex-nut driver, remove the nuts and washers securing the four BNC
connectors (1) on the rear panel.
8. Using the 5.0 mm driver, remove the hex screws (2) and washers that secure the A17 Baseband
Generator Interface to the rear panel.
9. Remove the A17 Baseband Generator Interface.
Replacement Procedure
Reverse the order of the removal procedure.
Torque all hex screws to 9 in-lbs.
Perform the post-repair adjustments and performance tests that pertain to this removal
procedure.

Table of Contents

Other manuals for Agilent Technologies E8257C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E8257C and is the answer not in the manual?

Agilent Technologies E8257C Specifications

General IconGeneral
BrandAgilent Technologies
ModelE8257C
CategoryPortable Generator
LanguageEnglish

Related product manuals