Assembly Replacement
Rear Panel (Standard)
9. Disconnect the following ribbon cables:
W64 from A17P2 (E8267C only)
W90 from A17P1 (E8267C only)
W20 from A20 SMI J1
W18 from A21 Rear-Panel Interface J3
W23 from A21 Rear-Panel Interface J2
10.If the signal generator has option UNR, disconnect the W48 flexible cable from A32 10 MHz
Crystal Oscillator (Option UNR).
11.Using the T-10 driver, remove the two screws (1) that secure the Rear Panel to the left and right
struts.
12. Using the T-10 driver, remove the four screws (2) that attach the rear panel to the frame.
13. Using the T-10 driver, remove the two screws (3) that secure the Rear Panel to the left and right
struts.
14.Using the needle-nose pliers, disconnect A22W1 from the power supply.
15.Using the T-10 driver, remove the screw securing the A22W2 to the A19 Power Supply.
16.Disconnect the B1 Fan cable from the A31 Motherboard P251.
17. Pull the rear panel away from the signal generator’s chassis.
Replacement Procedure
Reverse the order of the removal procedure.
Torque all T-10 screws to 9 in-lbs.
Torque all RF connectors to 8 in-lbs.
Perform the post-repair adjustments and performance tests that pertain to this removal
procedure.

Table of Contents

Other manuals for Agilent Technologies E8257C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E8257C and is the answer not in the manual?

Agilent Technologies E8257C Specifications

General IconGeneral
BrandAgilent Technologies
ModelE8257C
CategoryPortable Generator
LanguageEnglish

Related product manuals