2-28
Assembly Replacement
A11 Pulse/Analog Modulation Generator
A11 Pulse/Analog Modulation Generator
Tools Required
needle-nose pliers
Removal Procedure
Refer to Figure 2-12 for this procedure.
1. Disconnect the power cord.
2. Remove the covers from the signal generator. Refer to "Outer Instrument Cover" on page 2-5
and "Inner Instrument Cover" on page 2-7.
3. Position the signal generator with the front panel facing you.
4. Using the needle-nose pliers, disconnect the following flexible cables:
W1 from A11J401
W2 from A11J402
W3 from A11J403
W5 from A11J201
W17 from A11J205
W16 from A11J204
W7 from A11J202
W6 from A11J203
5. Lift the retention levers and disconnect the A11 Pulse/Analog Modulation Generator from the
A31 Motherboard.
6. Lift the A11 Pulse/Analog Modulation Generator out of the slot.
Replacement Procedure
Reverse the order of the removal procedure.
Perform the post-repair adjustments and performance tests that pertain to this removal
procedure.

Table of Contents

Other manuals for Agilent Technologies E8257C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E8257C and is the answer not in the manual?

Agilent Technologies E8257C Specifications

General IconGeneral
BrandAgilent Technologies
ModelE8257C
CategoryPortable Generator
LanguageEnglish

Related product manuals