2-66
Assembly Replacement
A25 20 GHz Highband Detector
A25 20 GHz Highband Detector
Tools Required
5/16” open ended wrench
needle-nose pliers
Removal Procedure
Refer to Figure 2-32 for this procedure.
1. Disconnect the power cord.
2. Remove the covers from the signal generator. Refer to "Outer Instrument Cover" on page 2-5
and "Inner Instrument Cover" on page 2-7.
3. Position the signal generator with the A31 Motherboard facing away from you and the RF deck
on top.
4. Using the 5/16” open ended wrench, disconnect the A25 20 GHz Highband Detector according to
the signal generator you have.
E8247C/57C: Disconnect the A25 Highband Detector from the A24 20 GHz Highband Coupler.
E8267C: Disconnect the A25 20 GHz Highband Detector from the A33 6dB PAD.
5. Using the needle-nose pliers, disconnect W82 from the A25 20 GHz Highband Detector.
6. Remove the A25 20 GHz Highband Detector.
Replacement Procedure
Reverse the order of the removal procedure.
Perform the post-repair adjustments and performance tests that pertain to this removal
procedure.

Table of Contents

Other manuals for Agilent Technologies E8257C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E8257C and is the answer not in the manual?

Agilent Technologies E8257C Specifications

General IconGeneral
BrandAgilent Technologies
ModelE8257C
CategoryPortable Generator
LanguageEnglish

Related product manuals