EasyManuals Logo

Agilent Technologies E8257C User Manual

Agilent Technologies E8257C
291 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #58 background imageLoading...
Page #58 background image
1-48
Troubleshooting
Troubleshooting Assembly–Level Problems
1001 10 MHz Test
1. Using an oscilloscope, check P41–2 for a 10 MHz signal greater than 1.12 V
pp
. If the signal is
present, replace the A7 Reference.
2. If the signal is not present, replace the internal 10 MHz standard.
1002 1 GHz Test
Replace the A7 Reference.