1-61
Troubleshooting
Troubleshooting Assembly–Level Problems
1608 VCO Tune Test
If the +32 V, ±15V, –5.2 V2, and + 5.2V supply voltages are good on the motherboard, use an
oscilloscope to check the 10 MHz signal on the A31 Motherboard at P161 pins 26 and 91. The
signal should have a period of 100 ns with an amplitude greater than 0.4 Vpp.
If the 10 MHz signal is good, replace the A14 Baseband Generator.
If the 10 MHz signal is bad, check it at the A7 Reference.
If the 10 MHz signal is bad at the A7 Reference, refer to A7 Reference troubleshooting.
If the 10 MHz signal is good at the A7 Reference, replace the A31 Motherboard.
If the +32 V, ± 15 V, and +5.2 V supply voltages are bad on the motherboard, check the origin of
the supply.
If the supply voltage is good at the origin, replace the A31 Motherboard.
If the supply voltage is bad at the origin, troubleshoot the A19 Power Supply.
If the –5.2 V2 supply voltage is bad on the motherboard, check the -7 V supply voltage at the A19
Power Supply.
If the -7 V supply voltage is good at the A19 Power Supply, replace the A31 Motherboard.
If the -7 V supply voltage is bad at the A19 Power Supply, troubleshoot the A19 Power supply.
1609 IQ DAC Output
Replace the A14 Baseband Generator
1610 PLL Test
Replace the A14 Baseband Generator

Table of Contents

Other manuals for Agilent Technologies E8257C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E8257C and is the answer not in the manual?

Agilent Technologies E8257C Specifications

General IconGeneral
BrandAgilent Technologies
ModelE8257C
CategoryPortable Generator
LanguageEnglish

Related product manuals