Damage hazards
!! DAMAGE HAZARD !! Do not use harsh abrasives, corrosive cleaning agents,
or metal brushes to clean any VSpin components or accessories. Do not use
any concentration of bleach (sodium hypoclorite). Do not allow cleaning agents
to contact any electrical or sensitive mechanical components.
!! DAMAGE HAZARD !! Plates occasionally break apart inside the VSpin, and
the plate fragments need to be removed so they do not interfere with moving
parts and potentially damage the VSpin.
!! DAMAGE HAZARD !! Do not tamper with or adjust the VSpin rotor mounting
screw, which fastens the rotor to the central motor axle. It is located on the
bottom of the rotor. Contact Automation Solutions Technical Support if you
think the rotor mounting screw requires maintenance. The VSpin is designed to
have a permanently attached rotor, unlike many manual centrifuges that can
have interchangeable rotors.
Safety labels
Warnings in the user documentation or on the device must be observed during
all phases of operation, service, and repair of this device. Failure to comply
with these precautions violates safety standards of design and the intended use
of the product. Agilent Technologies assumes no liability for the customer’s
failure to comply with these requirements. The following table lists the
common symbols. A description of the warning and information that will help
you avoid the safety hazard are provided in this guide.
Symbol Description
Indicates that you must read the accompanying
instruct
ions (for example, the safety guide) for more
information before proceeding.
Indicates hazardous voltages.
Indicates pinch, crush, and cut hazard.
Indicates laser hazard.
Indicates hot surface hazard.
Indicates protective conductor terminal, which is bonded
t
o conductive
parts of an equipment for safety purposes.
Indicates that you must not discard this electrical/
electronic product i
n domestic household waste.