Chapter 5: Maintenance and Troubleshooting
VSpin User Guide
Disconnecting the Air Supply
When to Use Disconnect the air supply before:
Performing maintenance on the VSpin
Cleaning the VSpin
Cleaning up broken plates
To disconnect the air supply:
1. Turn off the air supply so that there is no air pressure to the VSpin.
2. Push in the orange locking collar.
3. Gently pull on the air hose with your free hand until the air hose
comes out.
00080
VSpin
attach air hose
00080
Orange locking collar