73
Chapter 5: Maintenance and Troubleshooting
VSpin User Guide
Cleaning the VSpin
Overview To clean the VSpin, use soap, a mild detergent or an organic solvent
such as isopropyl alcohol (propan-2-ol) on a cloth. Wipe off the inner
shield surfaces, rotor and buckets. The buckets can be removed for
thorough cleaning and inspection.
Cleaning Hazards !! DAMAGE HAZARD !! Do not use harsh abrasives, corrosive
cleaning agents or metal brushes to clean any VSpin components
or accessories. Do not use any concentration of bleach (sodium
hypoclorite). Do not allow cleaning agents to contact any
electrical or sensitive mechanical components, such as the
central motor seal or the bucket pivot bearings.
!! DAMAGE HAZARD !! Trapped liquids may cause corrosion.
Do not let liquids contact:
❑ The electronic components
❑ Mechanical components inside the motor assembly
❑ The spindle support assembly
Cleaning Procedure !! INJURY HAZARD !! Disconnect the power, air and
communications cables before performing any maintenance or
cleaning.
To clean the VSpin:
1. Remove the buckets.
See “Removing a Bucket” on page 71.
This allows a more thorough cleaning and inspection of the VSpin
interior.
2. Clean the VSpin, using the guidelines stated above.
3. Replace the buckets.
See “To install or replace a bucket:” on page 31.