Chapter 5: Maintenance and Troubleshooting
VSpin User Guide
NmcNoOp() failed:
Communication likely failed
The VSpin may have become
disconnected during operation.
Check to make sure that the:
❑ Power cord is properly
plugged in to the VSpin
power entry and the wall
outlet
❑ Power switch is turned on
❑ Serial cable is properly
plugged into the VSpin and
the computer
Profile does not exist The VSpin profile does not exist. Create a new profile or select an
existing one.
Sensor failed to indicate bucket
unlock before the attempted
move
The sensor is not working
properly, the door, bucket or lock
is not in a state that allows a spin
to occur, or there is a bucket
imbalance.
Contact Velocity11.
Sensor failed to indicate door
lock before the attempted move
Sensor indicated imbalance
before the attempted move
ServoClearBits() failed Internal error. Contact Velocity11.
ServoLoadTraj() failed
ServoResetPos() failed
ServoSetGain() failed
ServoSetHoming() failed
ServoStopMotor() failed
Error What this means What to do