10.2.2 Measurement Calibration
All measurement calibrations are performed by adjusting the measurement reading up or
down using a calibration coefficient. The coefficient value ranges from -127 (adjust full-scale
downward, zero upward) to + 127 (adjust full-scale upward, zero downward). When used in
combination with the MEAS commands, an automated measurement calibration procedure
can be implemented by adjusting the coefficient and reading the new measurement value
iteratively until the delta between an external reference measurement device and the AC
source internal measurement is within the specified accuracy.
CALibrate:MEASure:CURRent[:FSCale] <NRf>
This command affects the calibration of the AC current measurement at full scale.
Command Syntax CALibrate:MEASure:CURRent[:FSCale] <NRf>
Parameters <NRf> (value range -127 to +127)
Examples CAL:MEAS:CURR -23
Query Syntax CALibrate:MEASure:CURRent?
Returned Parameters <NR1> (value range -127 to +127)
Related Commands CALibrate:MEASure:VOLTage
CALibrate:MEASure:CURRent:ZERO <NRf>
This command affects the calibration of the AC current measurement zero offset.
Command Syntax CALibrate:MEASure:CURRent:ZERO <NRf>
Parameters <NRf> (value range -127 to +127)
Examples CAL:MEAS:CURR:ZERO +12
Query Syntax CALibrate:MEASure:CURRent:ZERO?
Returned Parameters <NR1> (value range -127 to +127)
Related Commands CALibrate:MEASure:CURRent
CALibrate:MEASure:PCURrent[:FSCale] <NRf>
This command affects the calibration of the AC peak current measurement at full scale.
Command Syntax CALibrate:MEASure:PCURrent[:FSCale] <NRf>
Parameters <NRf> (value range -127 to +127)
Examples CAL:MEAS:PCUR -23
Query Syntax CALibrate:MEASure:PCURrent?
Returned Parameters <NR1> (value range -127 to +127)
Related Commands CALibrate:MEASure:VOLTage
CALibrate:MEASure:PCURrent:ZERO <NRf>
This command affects the calibration of the AC current measurement zero offset.
Command Syntax CALibrate:MEASure:PCURrent:ZERO <NRf>
Parameters <NRf> (value range -127 to +127)
Examples CAL:MEAS:PCUR:ZERO +12
Query Syntax CALibrate:MEASure:PCURrent:ZERO?
Returned Parameters <NR1> (value range -127 to +127)
Related Commands CALibrate:MEASure:CURRent