[SOURce:]FREQuency
This command sets the frequency of the output sinewave.
Command Syntax [SOURce:]FREQuency <NRf>
Parameters range specified by the LIM:FREQ:LOW and LIM:FREQ:HIGH
commands
Unit Hz (Hertz)
*RST Value defined by PON
Examples FREQ 50
Query Syntax FREQuency?
Returned Parameters <NR2>
Related Commands LIM:FREQ;LOW? LIM:FREQ:HIGH?
[SOURce:]VOLTage
This command programs the AC rms output voltage level of the power source. The
maximum voltage value allowed is determined by the selected voltage range. For 2003RP-
AV models configured for single voltage range, only one range is available.
Command Syntax [SOURce:]VOLTage[:LEVel] <NRf>
Parameters 0 to maximum voltage range specified by the VRAN command
Unit V (rms voltage)
*RST Value 0 volt
Examples VOLT 250 VOLT:LEV 25
Query Syntax [SOURce:]VOLTage[:LEVel]?
Returned Parameters <NR2>
Related Commands VOLT:RANGe