5-4 | Model 241CE II Hydrocarbon Dewpoint Analyzer
Propane Vapour Pressure
Vapour pressure data for propane from three sources
1,2,3
are tabulated in
Figures 5-1 and 5-2. The data in these figures are illustrated for compari-
son with the vapour pressures calculated from the equations listed under
References 2 and 3 (in Figures 5-1 and 5-2). At a given temperature, the
average of the results obtained from the two equations is taken as the
vapour pressure for propane. For completeness, the two equations are
reproduced here.
From Reference 2, the Pv is the vapour pressure of propane (mm Hg) and
t is the temperature (°C).
From Reference 3,
where Pv is the vapour pressure propane (mm Hg) and T is the absolute
temperature (K). Equation 2 was based upon 105 data points in the range
-128.9 to 96.8 °C and fitted the data with an average deviation of 0.46 per-
cent.
log
10
P
v
= 6.82973 –
813.2
248 + t
log
10
P
v
= 36.007 – – 11.666 log
10
T + 0.0085187T
1737.2
T

Table of Contents

Other manuals for Ametek 241CE II

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Ametek 241CE II and is the answer not in the manual?

Ametek 241CE II Specifications

General IconGeneral
BrandAmetek
Model241CE II
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals