2-2 | Model 241CE II Hydrocarbon Dewpoint Analyzer
Sensitivity
± 0.1 °C (± 0.2 °F)
This is for the stability of temperature measurement only.
Maximum Water Content
500 mg/m
3
(equivalent to a +5 °C water dewpoint at 200 PSIG)
Customer Connections
Analog Outputs
4 isolated 4–20 mA
DC
self-powered (standard) or loop powered (optional).
Self-Powered:
Configured as a 2-wire transmitter, compatible with a loop impedance of
0–1000 ohms.
Loop-Powered:
1500 Vrms isolated analog output configured as a 2-wire transmitter. An
external power supply (15–30 VDC) capable of supplying 25 mA per loop
is required (supplied by customer). When 30 VDC is supplied, the maxi-
mum loop impedance is 1000 ohms.
6 independent sets of SPDT Form C, dry (potential free) relay contacts,
24 VDC, 0.5 A Maximum
Digital Communication
Local Service Port: RS-232
Remote Service Port: RS-422/RS-485 (2-wire or 4-wire) using the optional
serial communications, Modicon Modbus
®
proto-
col, or using the optional System 200 Configurator
Software installed on a laptop computer.
Electrical Requirements
Power Consumption
< 275 W
Maximum Current: 3 A
Supply Voltage
120 VAC, 50/60 Hz, ±10 % max.
240 VAC, 50/60 Hz, ±10 % max.

Table of Contents

Other manuals for Ametek 241CE II

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Ametek 241CE II and is the answer not in the manual?

Ametek 241CE II Specifications

General IconGeneral
BrandAmetek
Model241CE II
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals