EasyManua.ls Logo

Ametek 241CE II

Ametek 241CE II
192 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Maintenance and Troubleshooting | 6-17
Filterblock Maintenance
The Membrane Filters in the Filterblock Assembly should be inspected
periodically for signs of contamination or damage.
Clean the Filterblock:
If alarms that indicate the optical paths show signs of contamination
are displayed, such as:
- f H2O Optics (SCode 32)
- f HC Optics (SCode 64)
If there is evidence of baseline deviations (see “Baseline Settings” in
the Preventive Maintenance Schedule in this chapter).
Every 6 months (minimum).
Each time you clean the Measuring Cell.
Replace parts in the Filterblock:
• Annually.
If, while cleaning the Filterblock, the Membrane Filters are found to
be damaged (are creased, have holes, or are transparent) or severely
contaminated (with liquids, oils, or waxes).
See Chapter 7 for
replacement part ordering
numbers.
RUN Mode
Stat\HistorySCode

Table of Contents

Other manuals for Ametek 241CE II

Related product manuals