3-2 | Model 241CE II Hydrocarbon Dewpoint Analyzer
Safety Considerations
Before installing and powering up the analyzer, review and follow all
safety information in this chapter and under “Personnel and Equip-
ment Safety Information” following the Table of Contents. This infor-
mation describes procedures to follow to avoid personal injury and/
or damage to the equipment. All regulatory agency and company
safety procedures for your jurisdiction must be followed.
Under normal operating conditions, lethal concentrations toxic gases
from the sample stream may be present within the sample system. The
sample system is defined as all components in the analyzer system
through which sample gas passes.
A breathing apparatus must be worn when installing/removing
equipment from the sample gas extraction point. The sample stream is
under positive pressure, and injury or death from inhaling toxic gases
in the sample stream could result from attempting to install/remove
equipment without the use of a breathing apparatus.
Pre-Installation Requirements
Storage Prior to Installation
If the analyzer is stored for any period of time prior to installation, store
the equipment in an environment where it is not subject to dripping or
splashing liquids, corrosive gases, high humidity, or excessive heat or cold.
Recommended storage conditions include:
Temperature: -20 °C–60 °C (-4 °F–140 °F)
Relative Humidity: maximum 95 %, non-condensing
Failure to comply with these storage conditions will void your warranty.
NOTE
!
WARNING