Service and Parts | 6-13
Furnace Replacement
Furnace Replacement
Use Figure 8-6:
1. Disconnect cell furnace wires.
2. Remove the Thermocouple assembly from the furnace.
3. Remove the Cell.
4. Loosen the swage nut at the top and bottom of the furnace. You may
need to remove the flow sensor to get the furnace assembly out.
5. Remove the Furnace assembly.
6. Install the new Furnace assembly in place.
7. Reconnect the Thermocouple, Flow Sensor and Cell.
8. Reconnect the wires on the sensor board.
Figure 8-6. Furnace Assembly