| ix
Safety Notes
WARNINGS, CAUTIONS, and NOTES contained in this manual emphasize
critical instructions as follows:
An operating procedure which, if not strictly observed, may result in
personal injury or environmental contamination.
An operating procedure which, if not strictly observed, may result in dam-
age to the equipment.
Important information that should not be overlooked.
Burn hazard. Hot surface. Do not touch, allow to cool before servicing.
Electrical Safety
Up to 5 kV may be present in the analyzer housings. Always shut down
power source(s) before performing maintenance or troubleshooting. Only a
qualified electrician should make electrical connections and ground checks.
Any use of the equipment in a manner not specified by the manufacturer may
impair the safety protection originally provided by the equipment.
Grounding
Instrument grounding is mandatory. Performance specifications and safety
protection are void if instrument is operated from an improperly grounded
power source.
Verify ground continuity of all equipment before applying power.