PC Congurator Software | A-25
Perform an Automatic Calibration
Scheduled Auto Calibration
1. Click on Conguration_AutoCalCong tab.
Figure24.Calibration_AutoCalCongurationscreen.
Calibration Gas Duration
Dene how long you want each
calibration gas to ow through the
sensor before switching the gases.
The RCU automatically switches
the span and zero calibration gases
based on the calibration gas duration
times you specify.
Be sure to allow enough time for the gases to stabilize when you dene these times.
Calibration Frequency in days
Calibration Start Date
Click on the calendar or click on PC
Date/Time.
Calibration Start Time Hours:Minutes:Seconds
Auto Calibration Enabled This must be checked.
Next Calibrations Date/Time
This date and time must be later than
the Calibration Date and Time.
Calibration gas values entered under the Configuration_Cal Gas
Values tab, are visible on the Set Cal Gas Values and Timers screen.
Verify the values set for calibration or edit, if necessary.