Perform a Manual Calibration
Figure15.Calibration_SetGasestoCalibratescreen.
1. Click the CALIBRATION button.
2. Select the gases to be calibrated, or choose the “All” option. Click the
NEXT button to continue.
Figure16.Calibration_SetCalGasValuesandTimersscreen.
3. Enter the calibration gas values for the gases you have selected.
Minimum and maximum values are noted on the screen aside of each
entry field.
- These gas values match the span and zero calibration gas cylinders
you use to calibrate your system.