5-10 | Thermox WDG-V / VC / VCM with Blow Back
Blow Back Box Temp Low Box temperature of Blow Back compart-
ment is below setpoint threshold.
Bad box heater(s), bad
box connection, bad RTD,
bad RTD connection
Low Sample Flow Low sample ow detected Loss of aspirator air
plugged inlet probe,
plugged sample system.
Active only after warm-up
time expires
Cell Failure (Open) Cell resistance exceeds the normal limit Bad Cell, Open Cell Con-
nection
Combustible Detector Open Comb detector is open or not connected Bad Detector, Bad elec-
tronics.
Only when congured as
WDG-VC
Methane Detector Open CH
4
detector is open or not connected Bad Detector, Bad elec-
tronics.
Only when congured as
WDG-VCM
Cell Mv Mismatch The cell mv measurement does not match
the real time cell meas
Bad A/D circuit (Bad elec-
tronics)
Analog Output 1 Error The analog output measurement does not
match the setting
Open 4-20mA loop or
failed output or bad read-
back circuit
Flow Sensor Failure The ow sensor failed (open) Bad ow sensor or circuit
Cell T/C Measurement Mis-
match
The Redundant temperature measurement
does not match the real time meas
Bad T/C or circuit

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Ametek Thermox WDG-VC and is the answer not in the manual?

Ametek Thermox WDG-VC Specifications

General IconGeneral
BrandAmetek
ModelThermox WDG-VC
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals