Chapter 6 Maintain the System
Regular Maintenance
105
Applied Biosystems StepOne
and StepOnePlus
Real-Time PCR Systems
Installation, Networking, and Maintenance Guide
Notes
Perform the
Calibration
PHYSICAL INJURY HAZARD. During instrument operation, the
temperature of the sample block(s) can exceed 100 ° C. If the instrument has been used
recently, keep your hands away until the sample block(s) reaches room temperature.
1. In the main menu of the StepOne
software, select Instrument
Instrument Maintenance Manager.
2. In the Instrument Maintenance Manager, click Dye in the navigation pane.
3. In the Dye screen, select System Dye Calibration.
4. Click Start Calibration.
5. In the Setup screen, select Plate 1 is loaded into the instrument.
6. Click Next.
7. In the Run screen, click Start Run .
3
6
5
4
1
2