ECS Charging System
Systems equipped with ECS will display the icon when batteries are
charging.
• The LED will flash green for a fixed period of 8 hours. After that period,
the device will launch another 8 hour charging period if the batteries are
still below 80%. Otherwise, the LED will become solid.
• If a function is activated during the charging process, the ECS icon
will disappear and the green LED will become solid. After the function
button is released, the device will attempt to reconnect to the ECS
system after approximately 20 seconds only if the battery charge
level is still below 80%.
If the ECS system is not supplying voltage due to a system fault or power
failure, the display will show actual battery level as for a non ECS system. The
device will attempt to reconnect to the ECS system every 15 minutes after the
fault being detected. The device will attempt to reconnect to the ECS system
only if the battery charge level is below 80%.
ECS Charging System
Systems equipped with ECS will display the icon when batteries are
charging.
• The LED will flash green for a fixed period of 8 hours. After that period,
the device will launch another 8 hour charging period if the batteries are
still below 80%. Otherwise, the LED will become solid.
• If a function is activated during the charging process, the ECS icon
will disappear and the green LED will become solid. After the function
button is released, the device will attempt to reconnect to the ECS
system after approximately 20 seconds only if the battery charge
level is still below 80%.
If the ECS system is not supplying voltage due to a system fault or power
failure, the display will show actual battery level as for a non ECS system. The
device will attempt to reconnect to the ECS system every 15 minutes after the
fault being detected. The device will attempt to reconnect to the ECS system
only if the battery charge level is below 80%.