Setup
DualWave Arthroscopy Pump User’s Guide
DFU-0212 Rev. 1 Page 30 of 54
and Cannula tube ends. Conversely, as the THRESHOLD value increases, so
does the likelihood of the DualWave not recognizing shaver activation. This
can result in failure of the DualWave to provide for aspiration through the
shaver blade and result in retention of heat within the blade.
• As the DELAY value decreases the DualWave will have less time to analyze the
information received through the Shaver Detect circuit. This action can result
in a false detection. Increasing the DELAY value should alleviate this
condition. As the DELAY value increases, the accuracy of the determination of
the shaver handpiece state will improve, but at the highest levels, the DELAY
time may result in a perceived lag before the DualWave responds to the shaver
activation.
• The DualWave’s Shaver Detect function has been verified to function properly
with the Arthrex Adapteur II Shaver Console and Arthrex Synergy Resection
Console. It is the responsibility of the user facility to evaluate third-party
shaver consoles to determine the correct operating parameters and any
potential hazards of utilizing the DualWave with an alternate shaver console.
4.7 How to Set Up the Synergy Heads-Up Display
1. First obtain the Synergy System Integration Cable Kit (AR-3200-1040).
2. Connect the Synergy System Integration Cable between the rear panel of the
DualWave [19] and the rear of the Synergy Imaging Console communication
connections labeled RS-232.
3. Turn on the DualWave and Synergy Imaging Console.
NOTE: Refer to IFU 950-0052-00 for more information.
4.8 How to Set Up Pump Tubing
NOTE: These instructions describe the procedure for setting up the AR-6410,
AR-6415, AR-6411, or AR-6420.
1. Remove the clip from the pump tubing and insert the pressure line
connector [e] of the pump tubing into the tubing sensor coupler [2].
This step must be completed first to ensure accurate pressure sensing.
2. Open the inflow door [12] completely. Allow the door to rest against
the stop. The roller mechanism is now exposed.
3. Place the green-collared section of the pump tubing [c] into the inflow
tubing track [1] indicated by the green mark.
4. Guide the tubing boot [d] over the rollers and insert the output side of
the tubing boot into the tubing OUT guide.
NOTE: The pump tubing is connected properly when the green connector [c] on
the pump tubing is aligned with the green mark on the front panel of the
console.
5. Close the inflow door [12].