Setup
DualWave Arthroscopy Pump User’s Guide
DFU-0212 Rev. 1 Page 33 of 54
4
4
.
.
1
1
3
3
.
.
4
4
Tubing Sensor Coupler
The pump motor automatically deactivates when the tubing is disconnected from
the pump. If the tubing is disconnected during a surgical procedure it must be
replaced by new tubing. Do not reconnect the tubing to the pump as it could lead
to unreliable pressure.
WARNING!
If the tubing is disconnected from the pump it must be replaced. Do not attempt to
reconnect the tubing to the pump as it could lead to unreliable pressure.
4.14 Shutdown Procedure
The AR-6480 can be safely shut down at any time by powering down the console.
All tubing accessories must be discarded as biohazardous waste.