Cleaning and Disinfecting
DualWave Arthroscopy Pump User’s Guide
DFU-0212 Rev. 1 Page 40 of 54
6.4 Tubing
The tubing is supplied pre-packaged sterile by EO sterilization. Do not resterilize.
Refer to the Instructions for Use package insert (DFU-0140) for
detailed tubing cleaning and sterilization instructions included with
each tubing set. Additional copies of this insert can be obtained
from the Arthrex website at www.arthrex.com
, or by contacting your
local Arthrex representative.
WARNING!
The extension, patient, and/or outflow tubing must be replaced before each new
patient and/or procedure.
Every extension or patient tubing set is supplied with a sterile connector cap for
the pump tubing set connection. Use this connector cap to cover the pump tubing
set connector after each surgical procedure to maintain sterility and ensure safe use
throughout the entire surgical day.