Maintenance
Regular and proper maintenance of your DualWave Arthroscopy Pump is the best way to
protect your investment and avoid non-warranty repairs.
Recommended care and handling of the DualWave Arthroscopy Pump includes proper
day-to-day operation, cleaning, and sterilization which are extremely important to ensure
safe and efficient operation. It is important to visually inspect the tubing, foot pedal,
remote control, cable, connectors, and display before each use.
Your authorized Arthrex service department is extremely knowledgeable about the
Arthrex Medical DualWave Arthroscopy Pump, tubing and/or foot pedal and remote
control and will provide a competent and efficient service. Any services and/or repairs
carried out by any unauthorized repair facility may result in reduced performance of the
instruments or instrument failure.
8.1 Periodic Maintenance
The product should be inspected prior to and after each use to ensure that the foot
pedal, remote control, cable, strain relief, overmold, connector contacts, and
display are not damaged or worn. If it becomes necessary to return the foot pedal
and/or remote control to Arthrex for service, please sterilize the remote control
before shipping. If fluid or particles splash on the display, clean with a microfiber
cloth by gently wiping in a circular motion.