Electromagnetic Emissions
DualWave Arthroscopy Pump User’s Guide
DFU-0212 Rev. 1 Page 48 of 54
1
1
3
3
.
.
0
0
Electromagnetic Emissions
Table 17 Guidance and Manufacturer's Declaration - Electromagnetic Emissions
The AR-6480 DualWave Arthroscopy Pump is intended for use in the electromagnetic
environment specified below. The customer or the user of the AR-6480 DualWave
Arthroscopy Pump should ensure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment guidance
RF emissions
CISPR 11
Group 1
The AR-6480 DualWave Arthroscopy Pump uses RF
energy only for its internal function. Therefore its RF
emissions are very low and not likely to cause any
interference with nearby electronic equipment.
RF emissions
CISPR 11
Class A
The AR-6480 DualWave Arthroscopy Pump is suitable
for use in all establishments other than domestic and
those directly connected to the public low-voltage power
supply network that supplies buildings used for
domestic purposes.
Harmonic
emissions
IEC 61000-3-2
Class A
Voltage
fluctuations/
flicker emissions
IEC 61000-3-3
Complies
Table 18 System Cables
Type
Use
Shielded
Ferrite
Maximum
Length
Power Cords
Supply line power to the
console
No No 3.048 m (10 ft)
Universal
Shaver Detect
Cord
Supply line power to the
shaver console from the
DualWave console
No No 2.438 m (8 ft)
Synergy System
Integration
Cable Kit
System integration cables No No 2.438 m (8 ft)

Table of Contents

Other manuals for Arthrex DualWave

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Arthrex DualWave and is the answer not in the manual?

Arthrex DualWave Specifications

General IconGeneral
BrandArthrex
ModelDualWave
CategoryMedical Equipment
LanguageEnglish

Related product manuals