Setup
DualWave Arthroscopy Pump User’s Guide
DFU-0212 Rev. 1 Page 32 of 54
4.11 How to Change the Language Setting
The AR-6480 supports English, French, German, Italian, and Spanish. The default
language is English. To change the language setting for operator display
messaging, follow these instructions.
1. Power ON the AC mains power switch [3] on the AR-6480.
2. Press the Menu button.
3. Press the Language button.
4. Select the desired language.
5. Press ok. The language is now stored in the memory.
4.12 How to Test the Power Supply Voltages
1. Performed automatically as part of the power-up sequence.
4.13 Safe Setup and Performance
4
4
.
.
1
1
3
3
.
.
1
1
Abnormal Operation
The AR-6480 employs a dual-pressure sensor design. Microcontroller-based
internal circuitry monitors the sensors, as well as other circuit parameters, to
ensure that the pump remains within normal operating limits. In the event of a
fault, the pump motor is automatically disabled and an error message is displayed
on the operator display [5]. See Table 3 for a complete list of operator display
messages and section 10.0 for troubleshooting information.
NOTE: If abnormal console operation cannot be corrected, disinfect the pump,
repackage in the original shipping materials, and return to Arthrex,
accompanied by a brief description of the malfunction. Prior to shipping, it
is necessary to obtain a return authorization number from Arthrex.
4
4
.
.
1
1
3
3
.
.
2
2
Overpressure Sensing
The sensing circuitry in the AR-6480 detects the pressure of the fluid in the tubing.
The overpressure alarm can be activated when the flow is abruptly interrupted or
the joint is suddenly positioned in a way which reduces the joint capsule volume
(e.g., bending the knee joint to the “Figure 4” position).
If an overpressure event occurs (300 mmHg), a warning message reading *Over
Pressure* will flash on the TPVD and an audible alarm will sound. The pump
motor is automatically disabled until the pressure returns to the set range.
To reduce the pressure in a joint, open an outflow and/or manipulate the joint to a
stress-free position.
4
4
.
.
1
1
3
3
.
.
3
3
Roller Housing
The pump motor automatically deactivates when the roller housing door is
opened. A locking mechanism prevents access to the rotating parts while the
device is operating.

Table of Contents

Other manuals for Arthrex DualWave

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Arthrex DualWave and is the answer not in the manual?

Arthrex DualWave Specifications

General IconGeneral
BrandArthrex
ModelDualWave
CategoryMedical Equipment
LanguageEnglish

Related product manuals