Electromagnetic Emissions
DualWave Arthroscopy Pump User’s Guide
DFU-0212 Rev. 1 Page 49 of 54
Table 19 Guidance and Manufacturer's Statement - Electromagnetic Immunity
The AR-6480 DualWave Arthroscopy Pump is intended for use in the electromagnetic environment
specified below. The customer or the user of the AR-6480 DualWave Arthroscopy Pump should
ensure that it is used in such an environment.
Electromagnetic environment -
guidance
discharge (ESD)
IEC 61000-4-2
± 8 kV air
± 8 kV air
Floors should be wood, concrete, or
ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30%.
transient/burst
IEC 61000-4-4
supply lines
± 1 kV for
input/output
supply lines
± 1 kV for
input/output lines
Mains power quality should be that
of a typical commercial or hospital
environment.
IEC 61000-4-5
line(s)
± 2 kV line(s) to
line(s)
± 2 kV line(s) to
Mains power quality should be that
of a typical commercial or hospital
environment.
short
interruptions,
and voltage
variations on
power supply
input lines
IEC 61000-4-11
(>95 % dip in Uτ)
for 0.5 cycle
40 % Uτ
(60 % dip in Uτ)
for 5 cycles
70 % Uτ
(30 % dip in Uτ)
for 25 cycles
<5 % Uτ
(>95 % dip in Uτ)
(>95 % dip in Uτ)
for 0.5 cycle
40 % Uτ
(60 % dip in Uτ)
for 5 cycles
70 % Uτ
(30 % dip in Uτ)
for 25 cycles
<5 % Uτ
(>95 % dip in Uτ)
Mains power quality should be that
of a typical commercial or hospital
environment. If the AR-6480
DualWave Arthroscopy Pump
requires continued operation during
power mains interruptions, it is
recommended that the AR-6480
DualWave Arthroscopy Pump be
powered from an uninterruptible
power supply.
(50/60 Hz)
magnetic field
IEC 61000-4-8
Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical
commercial or hospital environment.
Note: UT is the AC mains voltage prior to application of the test level.