Setup
DualWave Arthroscopy Pump User’s Guide
DFU-0212 Rev. 1 Page 27 of 54
4.6 Shaver Detect
The AR-6480 detects when a shaver handpiece has been activated. The Shaver
Detect feature operates differently, depending on if an AR-6430 Outflow Tube Set
is installed on the DualWave. Shaver Detect can be set up in three different ways.
NOTE: The Shaver Detect feature will function as described in sections 4.6.34.6.4.
4
4
.
.
6
6
.
.
1
1
Shaver Detect, Synergy System Connection
For correct operation of Shaver Detect when connecting with a Synergy System
Integration Cable (AR-3200-1040):
1. Ensure the Synergy System Integration Cable is properly connected
between the Synergy Resection Console and the DualWave.
2. Attach the shaver handpiece and footswitch (if used).
3. Turn on the Synergy Resection Console.
4. Turn on the DualWave.
NOTE: For use only with DualWave software version 1.8 or greater and Synergy
Resection Console version 2.3 or greater. Refer to IFU 950-0052-00 for
more information.
NOTE: It is recommended that the Shaver Detect Power Cable not be connected
when using the Synergy System Integration cable for Shaver Detection.
NOTE: The Shaver Detect feature will function as described in sections 4.6.3
4.6.4.
4
4
.
.
6
6
.
.
2
2
Shaver Detect, Shaver Power Cable
The Shaver Detect feature depends on the differential of current draw from the
shaver system when it is powered on in a static state and when the shaver
handpiece is activated.
For correct operation of Shaver Detect when connecting with the Shaver Interface
Cable:
1. Ensure the Shaver Detect Power Cable is properly connected between
the Shaver console and the DualWave.
2. Attach the shaver handpiece and footswitch (if used).
3. Turn on the Shaver Console.
4. After the Shaver Console sequence, turn on the DualWave.
NOTE: The Shaver Detect feature will function as described in sections 4.6.3
4.6.4.
NOTE: The Shaver Detect settings adjustments are described in section 4.6.5.

Table of Contents

Other manuals for Arthrex DualWave

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Arthrex DualWave and is the answer not in the manual?

Arthrex DualWave Specifications

General IconGeneral
BrandArthrex
ModelDualWave
CategoryMedical Equipment
LanguageEnglish

Related product manuals