EasyManuals Logo

Arthrex DualWave User Manual

Default Icon
60 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #47 background imageLoading...
Page #47 background image
Sterilization
DualWave Arthroscopy Pump User’s Guide
DFU-0212 Rev. 1 Page 41 of 54
7
7
.
.
0
0
Sterilization
Sterilization capabilities, cleaning, disinfecting, handling, and storage of instrumentation
are the responsibility of qualified facility and/or user personnel. Qualified personnel must
still properly clean and disinfect the instruments prior to sterilization.
Refer to the Instructions for Use package insert (DFU-0144) for detailed
remote control cleaning and sterilization instructions included with each
remote control. Additional copies of this insert can be obtained from the
Arthrex website at www. arthrex.com
, or by contacting your local Arthrex
representative.
After sterilization in the autoclave, let the accessory device cool down
slowly. NEVER use cold water to cool the remote control. This will damage
the electronic components and seals.
Liquid on the cable connector of the accessory device can damage the
device. Before connecting the cable, ensure the receptacles are clean and
dry.
WARNING!
After autoclaving, the accessory devices are VERY HOT. Handle with care to avoid burns.
7.1 Transmissible Spongiform Encephalopathy Agents
It is outside the scope of this document to describe in detail the precautions that should be
taken for Transmissible Spongiform Encephalopathy (TSE) Agents.
The agents for transmission of Creutzfeldt-Jakob disease are believed to be resistant to
normal disinfection and sterilization processes. Therefore, the normal processing methods
of decontamination and sterilization as described above may not be appropriate where
CJD transmission is a risk.
In general, the tissues that come into contact with orthopedic surgical instruments are
those of low TSE infectivity. However, take particular precautions when handling
instruments that have been used on known, suspected, or at-risk patients.
References:
ANSI/AAMI ST79: Good Hospital Practice: Steam sterilization and sterility assurance.

Table of Contents

Other manuals for Arthrex DualWave

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Arthrex DualWave and is the answer not in the manual?

Arthrex DualWave Specifications

General IconGeneral
BrandArthrex
ModelDualWave
CategoryMedical Equipment
LanguageEnglish

Related product manuals