129
in the Interlock state and will not be able to generate output voltage or current.
LF?
Lists the file name of the file loaded into active memory (RAM).
LF <file number>?
List the file name of any saved file within the directory structure. The file number is
given to the file when the file is save into non-volatile memory.
LP?
Lists the prompt that is created for the individual step that has been selected within
active memory (RAM).
LP <step number>?
Lists the prompt that has been created for of one of the steps of the file within active
memory (RAM).
FT?
Returns the total quantity of files stored in memory.
ST?
Returns the total quantity of step created in the file.
LS?
Lists all the Parameters for the individual step that is currently selected.
The response will be formatted as follows; <step, test, p1, p2, p3…> Where <step> is
the step number, <test> is the test type and <p1,p2> etc., indicates the parameters of
the test. Please refer to the ADD command for a list of parameters for each test type.
LS <step number>?
Lists all the Parameters for the individual step indicated by step number = 1-30.
The response will be formatted as follows; <step, test, p1, p2, p3…> Where <step> is
the step number, <test> is the test type and <p1,p2> etc., indicates the parameters of
the test. Please refer to the ADD command for a list of parameters for each test type.
LS2?
Lists all the Parameters for the individual step that is currently selected.
The response will be formatted as follows; <step, test, p1, p2, p3…> Where <step> is
the step number, <test> is the test type and <p1,p2> etc., indicates the parameters of
the test. Please refer to the ADD command for a list of parameters for each test type.
LS2 <step number>?
Lists all the Parameters for the individual step indicated by step number = 1-30.
The response will be formatted as follows; <step, test, p1, p2, p3…> Where <step> is
the step number, <test> is the test type and <p1,p2> etc., indicates the parameters of
the test. Please refer to the ADD command for a list of parameters for each test type.
RS?

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Associated Research OMNIA II 8207 and is the answer not in the manual?

Associated Research OMNIA II 8207 Specifications

General IconGeneral
BrandAssociated Research
ModelOMNIA II 8207
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals