Periodic Maintenance AVL 4000
1-2 Service Manual
1.2 Maintenance Plan
To ensure that the tester always functions properly, maintenance must be carried out at
regular intervals.
Maintenance work
Daily
Weekly
Six-monthly
(minor maintenance)
One a year
(major maintenance)
Comment
Entry in main-
tenance log book
Leak test

Automatically required
by tester every day
Filter replacement
Pre-filter on probe

Particulate filter

Condensate filter

Activated carbon filter
Replace in good time
when soiled by HC
vapour (e.g. petrol
vapour).
Cleaning
Probe and hose

see Operating Manual
Condensate pump
(cleaning kit)

see Operating Manual
Calibration with calibrating
gas
as per statutory requirements
Replacing O
2
sensor
Replacement is
requested by tester
Replacing NO sensor Replacement is
requested by tester
Visual check of outside

Check using software
Service engineer
Software update (as required)
Service engineer
Open tester, visual inspection
of PCBs, clean, close housing,
seal with service seal
Service engineer