4000 Opacity Measuring Chamber Type A
Service Manual 6-15
6.7 Changeover Valve with Geared Motor
Dismantle the opacity measuring chamber as described in section 6.4 Dismantling the
Opacity Measuring Chamber.
The changeover valve is located in half-shell 1 (exhaust gas outlet side).
Simply pull the valve out of the half-shell by its aluminium housing.
Disconnect the cable to the geared motor.
The connector is held in place by barbs (on the side of the plug socket) to prevent
it from being pulled out inadvertently.
Release the barbs by pressing inwards with a small screwdriver.
Valve block BO1136
Fig. 6-24
Valve housing
YM2685
Sliding disc
DL0092
Valve bonnet 2
YM2687
Liner
DL0091
Valve vane
YM2688
Setscrew
DS0988
3 countersunk screws M3x16
DS0371
Geared motor 12 V DC
EI0189
Liner
DL0091
Valve bonnet 1
YM2686
Sliding disc
DL0092
Straight pin
DB0102