Periodic Maintenance AVL 4000
1-8 Service Manual
1.4 Major Maintenance
1.4.1 AVL DiGas 4000
Evaluation Unit
Check using software
Raw data from 4-gas measuring instruments unit OK?
O
2
sensor is OK?
NO sensor is OK?
Open evaluation unit (remove cover)
Visual inspection, clean pcb
If necessary update setpoint data and software
Close housing, seal with service seal
Enter in maintenance log book
Cleaning
Replace particulate filter
(see section 5.3.3 Particulate Filter)
Check O-ring paper filter
Visual inspection probe and hose
Clean condensate pump
(using cleaning kit)
Function test
Leak test
Gas calibration