EasyManua.ls Logo

AVL DiGas 4000 - Service Mode; Changing O and no Sensor; Gas Calibration; 5-Gas

Default Icon
182 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Evaluation Unit AVL 4000
4-54 Service Manual
4.25 Service Mode
4.25.1 Changing O
2
and NO Sensor
see Section 5.6 or Operating Manual
4.25.2 Gas Calibration
see Operating Manual
4.25.3 4-/5-Gas
Menu for status assessment and error diagnosis of the 4-/5-gas measuring instrument
unit
Remember that access to this measurement screen may be protected depending on the
software version.
General notes
Compare the current raw data with that from the last service (service printout).
Error messages in fields 3 and 4 suggest a fault in the hardware.
Check the pneumatic components unit, cables etc. before sending the measuring
instrument unit back to the works.
Observe the initial response of the gain factors (field 9).
If the target values are not reached, this suggests a fault in the hardware.
Remember that the instrument must be switched on for at
least 30 minutes before the calibration.
If the measuring instrument unit develops a fault and has to
be returned to the works, always include a "service printout"
to assist the error diagnosis.

Table of Contents

Related product manuals